FDA recall Z-1808-2020

Uromedica Inc. · Class II · device

Product

ProACT Implantation Instrument Set, is a reusable, steam sterilizable surgical instrument set consisting of a U-Channel Sheath through which the Pro ACT balloon implant is introduced.

Reason for recall

Potential issue with the manufacture and inspection of U-channel sheaths in Implantation Instruments Sets. Affected U-channel sheaths may be more likely to cause damage to ProACT implant devices when passes through the sheaths.

Distribution

Nationwide distribution to the following states: CA, GA, FL, MI, and VA.

Key facts

Status
Terminated
Initiation date
2019-04-10
Report date
2020-05-06
Termination date
2020-05-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
Plymouth, MN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1808-2020