FDA recall Z-1809-2020

Philips North America, LLC · Class II · device

Product

Philips Azurion 7 M20 -XperGuide Software hosted in Interventional Workspot 1.5 when used with the Azurion R2.0 system Interventional fluoroscopic x-ray system System code: 722079

Reason for recall

When a user acquires XperCT scan on an Azurion 2.0 system, enters the XperGuide guidance step and moves the L-arm away from the initial scan position before starting the live guidance, a warning message directs the user to move the L-arm stand back to the initial XperCT scan position. Although the software generates this message, it does not prevent the use of live guidance if the L-arm stand is not repositioned. Using live guidance with a mispositioned L-arm can result in the display of an incorrect overlay and needle path.

Distribution

Worldwide distribution. US states of FL, ID, IL, IN, MA, and OH, Canada, Australia, Austria, Belgium, Germany, and Netherlands New Zealand, South Africa, Switzerland, Thailand

Key facts

Status
Terminated
Initiation date
2020-04-03
Report date
2020-05-06
Termination date
2023-02-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
Andover, MA, United States

Primary source