Dimension Magnesium (MG) Flex reagent cartridge Siemens Material Number (SMN)/REF (Catalog Number):10444963/DF57
Dimension Magnesium (MG) Flex reagent cartridge lots may exhibit imprecision for Quality Control (QC) and produce Abnormal Reaction flags. The issue is not always detected by QC and erroneous results may be produced in the absence of an Abnormal Reaction flag. Based on customer data, imprecision leading to erroneous, unflagged patient results with a bias of -15 to -59% may occur
Nationwided Foreign: Afghanistan¿¿¿¿ Albania¿¿¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Bulgaria¿¿¿¿¿¿¿ Canada Croatia¿¿¿¿¿¿¿¿ Czech¿Republic¿ Egypt¿¿¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Hungary¿¿¿¿¿¿¿¿ Israel¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Jordan¿¿¿¿¿¿¿¿¿ Kazakhstan¿¿¿¿¿ Kenya¿¿¿¿¿¿¿¿¿¿ Kuwait¿¿¿¿¿¿¿¿¿ Latvia¿¿¿¿¿¿¿¿¿ Lebanon¿¿¿¿¿¿¿¿ Macedonia¿¿¿¿¿¿ Mali¿¿¿¿¿¿¿¿¿¿¿ Malta¿¿¿¿¿¿¿¿¿¿ Montenegro¿¿¿¿¿ Morocco¿¿¿¿¿¿¿¿ Nepal¿¿¿¿¿¿¿¿¿¿ Netherlands¿¿¿¿ Pakistan¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Qatar¿¿¿¿¿¿¿¿¿¿ Romania¿¿¿¿¿¿¿¿ Russian¿Fed.¿¿¿ Saudi¿Arabia¿¿¿ Senegal¿¿¿¿¿¿¿¿ Serbia¿¿¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Slovenia¿¿¿¿¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Sudan¿¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ U.A.E.¿¿¿¿¿¿¿¿¿ Uganda¿¿¿¿¿¿¿¿¿ Ukraine¿¿¿¿¿¿¿¿ United¿Kingdom¿ Uzbekistan¿¿¿¿¿