FDA recall Z-1818-2020

WELCH ALLYN, INC/MORTARA · Class II · device

Product

Q-Tel. The Q-Tel RMS system is a computer-based cardiac and pulmonary rehabilitation data acquisition and editing system.

Reason for recall

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Distribution

US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.

Key facts

Status
Terminated
Initiation date
2020-03-18
Report date
2020-05-06
Termination date
2026-02-09
Voluntary/Mandated
Voluntary: Firm initiated
Location
Milwaukee, WI, United States

Primary source