FDA recall Z-1827-2020

Capso Vision, Inc. · Class II · device

Product

CapsoCAM Plus, UDI: 00867770000209

Reason for recall

Capsule Endoscopy System: a defect present in the capsule window may be stressed through the production process and in the final package, which could cause the housing of the capsule to crack and leak when ingested. Fluid from the patient could leak in to the capsule, which may damage components inside the capsule and the patient data may not be retrieved; the patient may need to repeat the exam.

Distribution

U.S.: CA, TX, FL, VA, PA, SC, HI, DC, NJ, UT, LA. Foreign: FR, AR, IT, UK, BE, TC, AU, CO, GM, GR, SP, SI, HR, MO, SW, EZ.

Key facts

Status
Terminated
Initiation date
2019-10-18
Report date
2020-05-06
Termination date
2021-01-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Saratoga, CA, United States

Primary source