EGIA 60 ARTICULATING EXTRA THICK SULU, Item code EGIA60AXT
The device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the potential for infection and/or sepsis.
Nationwide domestic distribution, worldwide foreign distribution.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1828-2019