FDA recall Z-1831-2021

Normand-Info S.A.S.U. · Class II · device

Product

Normand Remisol Advance Data Manager is a software device that interfaces between laboratory information systems (LIS) and laboratory instruments.

Reason for recall

Instructions For Use for software versions 1.5 to 1.9 did not provide sufficient instructions regarding the proper programming and testing of rules. Therefore, rules written by users that were not validated or incorrectly validated may lead to erroneous results.

Distribution

Worldwide distribution - US Nationwide distribution in the states of Alabama, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Indiana, Iowa, Kansas, Louisiana, Maine, Maryland, Massachusetts, Michigan, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, Pennsylvania, Rhode Island, South Carolina, Texas, West Virginia and Wyoming. The countries of Australia, Austria, Belgium, Brunei, Canada, China, Colombia, Congo, Czech Republic, Denmark, Ethiopia, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Lithuania, Malaysia, Mayotte, Morocco, Netherlands, Netherlands Antillean, New Zealand, Norway, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uganda, UAE, UK, and Zimbabwe.

Key facts

Status
Terminated
Initiation date
2021-03-22
Report date
2021-06-16
Termination date
2022-12-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
Arras CEDEX, N/A, France