FDA recall Z-1834-2022

Cook Incorporated · Class II · device

Product

Fixed Core Wire Guide (Straight), Guidewire Reference Part Number/GPN: C-SF-18-30 G02261 C-SF-25-40 G01937 TSF-18-260 G00590 C-SF-25-50 G01938 TSF-35-180 G00652 C-SF-18-40 G02255 TSF-25-145 G00617 TSF-32-145 G00638 C-SF-15-50 G02274 TSF-18-145 G00587 TSF-18-50 G00593 TSF-35-145 G00650 TSF-21-145 G00602 THSF-35-180 G01237 TSF-35-80 G00664 C-SF-25-30 G02163 C-SF-21-30 G02351 THSF-35-80 G02408

Reason for recall

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

Distribution

Domestic distribution nationwide. Foreign distribution worldwide.

Key facts

Status
Ongoing
Initiation date
2022-08-16
Report date
2022-10-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bloomington, IN, United States