FDA recall Z-1837-2022

Cook Incorporated · Class II · device

Product

Roadrunner PC Hydrophilic Wire Guide, Hydrophilic Guidewire Reference Part Number/GPN: RPC-035145-0-5 G34131 RPC-035145 G18154 RPC-035145-5 G34132 RPC-038145-0-5 G34129 RFSPC-038145-0 G17539 RFSPC-035145 G17540 RPC-038145-0 G18155 RPC-035145-0 G18153 RFSPC-035145-0 G17542

Reason for recall

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

Distribution

Domestic distribution nationwide. Foreign distribution worldwide.

Key facts

Status
Ongoing
Initiation date
2022-08-16
Report date
2022-10-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bloomington, IN, United States