Zenith Alpha Abdominal Endovascular Graft, Reference Part Numbers: ZIMB-22-108, ZIMB-22-118, ZIMB-22-128, ZIMB-22-70, ZIMB-22-84, ZIMB-22-98, ZIMB-24-108, ZIMB-24-118, ZIMB-24-128, ZIMB-24-70, ZIMB-24-84, ZIMB-24-98, ZIMB-26-108, ZIMB-26-118, ZIMB-26-128, ZIMB-26-70, ZIMB-26-84, ZIMB-26-98, ZIMB-28-108, ZIMB-28-118, ZIMB-28-128, ZIMB-28-70, ZIMB-28-84, ZIMB-28-98, ZIMB-30-108, ZIMB-30-118, ZIMB-30-128, ZIMB-30-70, ZIMB-30-84, ZIMB-30-98, ZIMB-32-108, ZIMB-32-118, ZIMB-32-128, ZIMB-32-70, ZIMB-32-84, ZIMB-32-98, ZIMB-36-108, ZIMB-36-118, ZIMB-36-128, ZIMB-36-70, ZIMB-36-84, ZIMB-36-98
may contain a damaged gray safety lock knob, which could potentially result in difficulty or an inability to fully deploy the graft via the standard or troubleshooting method provided in the Instructions for Use (IFU). Potential adverse events that may occur if an affected product is used include a prolonged procedure and open surgical intervention.
No US distribution. Product distributed to the following foreign countries: Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile Columbia, Costa Rica, Czech Republic, Denmark, Finland, France, Germany, Greece, Ireland, Israel, Italy, Japan, Morocco, Netherlands, Norway, Poland, Portugal, Reunion, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, UK