FDA recall Z-1844-2024

Philips North America Llc · Class II · device

Product

Evolution upgrade 3.0T Model Number (REF): (1)782117; (2) 782143

Reason for recall

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

Distribution

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.

Key facts

Status
Ongoing
Initiation date
2024-05-03
Report date
2024-05-29
Voluntary/Mandated
Voluntary: Firm initiated