FDA recall Z-1846-2020

Applied Medical Resources Corp · Class I · device

Product

Python Embolectomy Catheters, Sterile EO, CE, Ref numbers/UDI: A4E01/(01)00607915113407; A4E02/(01)00607915113414; A4E03/(01)00607915113421; A4E04/(01)00607915113438; A4E05/(01)00607915113445; A4E06/(01)00607915113452; A4E08/(01)00607915113476; A4E09/(01)00607915113483

Reason for recall

The firm has received reports of tip separation during usage of the catheters, which may expose patients to unintended dislodgement of the tip from the body of the catheter.

Distribution

Worldwide distribution - United States: Nationwide OUS: Australia, Belgium, Canada, Czech Republic, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Netherlands, Poland, Saudi Arabia, South Africa, Spain, Taiwan, United Kingdom

Key facts

Status
Terminated
Initiation date
2019-10-24
Report date
2020-05-13
Termination date
2023-06-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
Rancho Santa Margarita, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1846-2020