# FDA recall Z-1866-2019

> **Thermo Fisher Scientific (Asheville) LLC** · Class II · device recall initiated 2019-04-02.

## Product

Fisher Scientific Isotemp -General Purpose Water Bath:   lsotemp 10L GP Bath  Model Number: FSGPD10

## Reason for recall

Product may have been assembled incorrectly,the  assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.

## Distribution

CA, GA, KY, PA  Foreign: Canada, China, Germany, Japan, New Zealand, Australia

## Key facts

- **Recall number:** Z-1866-2019
- **Recalling firm:** Thermo Fisher Scientific (Asheville) LLC
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2019-04-02
- **Report date:** 2019-07-03
- **Termination date:** 2020-07-23

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Marietta, OH, United States