FDA recall Z-1874-2019

Hamilton Medical AG · Class I · device

Product

HAMILTON-G5, with software versions less than or equal to 2.60 The HAMILTON-G5/S1 ventilators are designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The device is intended for use in the hospital and institutional environment where health care professionals provide patient care.

Reason for recall

New software version for affected ventilators reduces the probability of the ventilator entering an ambient state, in which the inspiratory channel and expiratory valves are opened, letting the patient breathe room air unassisted. When the ventilator enters the ambient state, alternative ventilation must be provided immediately.

Distribution

Worldwide Distribution- US (Nationwide Distribution): MI, WI, OR, MO, TX, NV, PA, SC, HI, MN, ND, NC, AZ, FL, CA, MA, GA, NY, VA, WA, IN, CO, LA, IL, OH, TN, SD, OK, MI, AR, C, AK, WV, NJ, IA, NE, KS, MT, NH, AL, MD, KY, CT, and RI. And (International Distribution): AT, BE, BG, CH, CZ, DE, DK, EE, ES, FI, FR, GB, GR, HU, IT, MT, NL. NO, PL.PT, RO, SE, SI, SK, TR, HR. AGO, ARE, ARM, AUS, AZE, BGD, BHR , BLR, BOL, BRA, CAN, CHL, CHN, COL, CRI, DOM, DZA, ECU, EGY, EST, HKG, IDN, IND, IRQ, ISR, JOR, JPN, KAZ, KOR, KWT, LBN, LBY, LKA, LVA, MAR, MDA, MDV, MEX, MNG, MYS, NPL, NZL, OMN, PAK, PAN, PER, PHL, PRI, PSE, QAT, RUS, SAU, SGP, SUR, SYR, THA, TUN, TWN, UKR, URY, USA, VEN, VNM, YEM, and ZAF.

Key facts

Status
Terminated
Initiation date
2019-03-25
Report date
2019-07-17
Termination date
2024-06-03
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bonaduz, Switzerland

Primary source