# FDA recall Z-1875-2019

> **GE Healthcare, LLC** · Class III · device recall initiated 2019-05-10.

## Product

ECG Trunk Cables:   REF/Catalog Number	Description  2106305-001	ECG Trunk Cable, 3/5-lead, AHA, 3.6 m/12 ft.  2106305-002	ECG Trunk Cable, 3/5-lead, AHA, 1.2 m/4 ft.  2106305-003	ECG Trunk Cable, 3/5-lead, IEC, 3.6 m/12 ft.  2106305-004	ECG Trunk Cable, 3/5-lead, IEC, 1.2 m/4 ft.  2106306-001	ECG Trunk Cable, Neonatal, DIN 3-lead, AHA, 3.6 m/12 ft.  2106306-002	ECG Trunk Cable, Neonatal, DIN 3-lead, AHA, 1.2 m/4 ft.  2106306-003	ECG Trunk Cable, Neonatal, DIN 3-lead, IEC, 3.6 m/12 ft.  2106306-004	ECG Trunk Cable, Neonatal, DIN 3-lead, IEC, 1.2 m/4 ft.  2106307-001	ECG Trunk Cable, 6-lead, AHA, 3.6 m/12 ft.  2106307-002	ECG Trunk Cable, 6-lead, AHA, 1.2 m/4 ft.  2106307-003	ECG Trunk Cable, 6-lead, IEC, 3.6 m/12 ft.  2106307-004	ECG Trunk Cable, 6-lead, IEC, 1.2 m/4 ft.  2106308-001	ECG Trunk Cable, 12-lead, AHA, 3.6 m/12 ft.  2106308-002	ECG Trunk Cable, 12-lead, AHA, 1.2 m/4 ft.  2106308-003	ECG Trunk Cable, 12-lead, IEC, 3.6 m/12 ft.  2106308-004	ECG Trunk Cable, 12-lead, IEC, 1.2 m

## Reason for recall

ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

## Distribution

Worldwide Distribution: US (nationwide)  including Puerto Rico, and countries of: Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malta, Mauritius, Mexico, Moldova, Morocco, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Palestine (state of), Panama, Poland, Qatar, Republic of Serbia, Romania, Russia, Saudia Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad and Tobago, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Ur

## Key facts

- **Recall number:** Z-1875-2019
- **Recalling firm:** GE Healthcare, LLC
- **Classification:** Class III
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2019-05-10
- **Report date:** 2019-07-03
- **Termination date:** 2024-05-24

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Waukesha, WI, United States