# FDA recall Z-1877-2024

> **Getinge Usa Sales Inc** · Class I · device recall initiated 2024-05-01.

## Product

Vaporizer Sevoflurane Maquet Filling, Product code/REF number (6682282), Component of: Flow-i C20, Flow-i C30,Flow-i C40, Flow-c, Flow-e.

## Reason for recall

Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure.  Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.

## Distribution

Domestic: CA, CO, IN, MO, OR, VA, WA.   International: Aruba, Australia, Austria, Belgium, Bolivia, Bosnia-Herzegovina, Brazil, Bulgaria, Chile, China, Costa Rica, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macau, Macedonia, Malaysia, Mexico, Moldavia, Morocco, Mozambique, Namibia, Netherlands, Nicaragua, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Vietnam Please note that a list of foreign consignees was not provided by the manufacturer and it is not available at this time.

## Key facts

- **Recall number:** Z-1877-2024
- **Recalling firm:** Getinge Usa Sales Inc
- **Classification:** Class I
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-05-01
- **Report date:** 2024-06-12

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Wayne, NJ, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1877-2024

## Citation

> AI Analytics. FDA recall Z-1877-2024. Retrieved 2026-07-26 from https://api.ai-analytics.org/recall/Z-1877-2024. Source: US FDA. Licensed CC0.

---