FDA recall Z-1884-2025

Onkos Surgical, Inc. · Class II · device

Product

Male-Female Midsection with the following sizes and model numbers. a. Male-Female Midsection, 40mm, 1 EACH; Model Number: 25001040E b. Male-Female Midsection, 50mm, 1 EACH; Model Number: 25001050E c. Male-Female Midsection, 60mm, 1 EACH; Model Number: 25001060E d. Male-Female Midsection, 70mm, 1 EACH; Model Number: e. Male-Female Midsection, 90mm, 1 EACH; Model Number: 25001090E f. Male-Female Midsection, 110mm, 1 EACH; Model Number: 25001110E g. Male-Female Midsection, 140mm, 1 EACH. Model Number: 25001140E Component of ELEOS SEGMENTAL LIMB SALVAGE SYSTEM

Reason for recall

Potential for breach of Tyvek seals.

Distribution

Product was shipped from Onkos' distribution in Olive Branch, MS to consignees in the following states: GA, TX, WA, LA, NJ, FL, AZ, MD, MS, KS, IL, WI, KY, PA, NC, NE, OH, HI.

Key facts

Status
Ongoing
Initiation date
2023-09-01
Report date
2025-06-04
Voluntary/Mandated
Voluntary: Firm initiated
Location
Parsippany, NJ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1884-2025