FDA recall Z-1891-2020

Intuitive Surgical, Inc. · Class II · device

Product

daVinci XI Sureform 60 Reload, 60 mm black 6-Row 4.6 mm, Ref: 48360T-08 , UDI: 00886874113851, Rx only, SterileEO - Product Usage: are intended to be used with da Vinci Surgical Systems for resection, transection, and, or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery.

Reason for recall

The firm became aware of a manufacturing variation in which black and green reloads in select Stapler Reloads may not deploy individual adjacent staples, which could result in an incomplete staple line. Failure of a incomplete staple like could potentially lead to an air leak or an anastomotic leak which may require an additional procedural intervention.

Distribution

Worldwide distribution - US Nationwide including in the states of AL, AK, AZ, AR, CA, CO, CT, DE, D.C., FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, WI and the countries of Australia, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Luxembourg, Netherlands, Norway, South Korea, Spain, Switzerland, United Kingdom.

Key facts

Status
Ongoing
Initiation date
2019-07-01
Report date
2020-05-13
Voluntary/Mandated
Voluntary: Firm initiated
Location
Sunnyvale, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1891-2020