FDA recall Z-1898-2024

FUJIFILM Healthcare Americas Corporation · Class II · device

Product

FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. Model/Catalog Number: 566-16130-23 566-16130-33

Reason for recall

Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury

Distribution

US Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.

Key facts

Status
Ongoing
Initiation date
2024-04-11
Report date
2024-06-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Lexington, MA, United States

Primary source