FDA recall Z-1902-2018

Smith & Nephew, Inc. · Class II · device

Product

smith&nephew 5o (5 degree) GENESIS(R) II VALGUS BUSHING, REF CATALOG NUMBER 71440014 The GENESIS II Knee System Components are indicated for rheumatoid arthritis; post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity levels are compatible with an adequate long-term result; failed osteotomies, unicompartmental replacement, or total knee replacement. The constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e., medial collateral and/or lateral collateral ligament) are absent or incompetent.

Reason for recall

A single lot of GENESIS II FEMORAL FIVE DEGREE VALGUS BUSHING due to a laser mark labeling error. The left valgus bushing was incorrectly laser marked as right and vice versa.

Distribution

US Distribution to the states of : NC, IN, CO MI, PA, NY, MA, CA, GA

Key facts

Status
Terminated
Initiation date
2018-04-09
Report date
2018-05-30
Termination date
2019-02-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Memphis, TN, United States

Primary source