FDA recall Z-1902-2019

Siemens Medical Solutions USA, Inc · Class II · device

Product

Siemens Artis Q.zen floor----Interventional Fluoroscopic X-Ray System Model Number: 10848353

Reason for recall

Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel

Distribution

Nationwide

Key facts

Status
Terminated
Initiation date
2019-06-06
Report date
2019-07-10
Termination date
2021-02-09
Voluntary/Mandated
Voluntary: Firm initiated
Location
Malvern, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1902-2019