# FDA recall Z-1909-2024

> **Stryker Neurovascular** · Class II · device recall initiated 2024-04-23.

## Product

Synchro .014"guide wire, 35 cm tip, Catalog: M00313010  Synchro .014 guide wire, 300 cm length, Catalog: M00313310  SYNCHRO .010 WIRE SOFT STRAIGHT 200cm, Catalog: M00316310  SYNCHRO .010 WIRE SOFT STRAIGHT 300cm, Catalog: M00316330

## Reason for recall

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.

## Distribution

Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, SAR China, Hungary, India, Ireland, Israel, Italy, Japan, Korea (South), Lithuania, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom.

## Key facts

- **Recall number:** Z-1909-2024
- **Recalling firm:** Stryker Neurovascular
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-04-23
- **Report date:** 2024-06-05

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Fremont, CA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1909-2024

## Citation

> AI Analytics. FDA recall Z-1909-2024. Retrieved 2026-07-04 from https://api.ai-analytics.org/recall/Z-1909-2024. Source: US FDA. Licensed CC0.

---