FDA recall Z-1915-2020

ICU Medical, Inc. · Class II · device

Product

REF: MC330424, 145" (368 cm) Appx 19.8 ml, 10 Drop Primary Set w/4 MicroClave Clear, Remv 3-Way Stopcock, Rotating Luer, 1 Ext, Drop-In Ext Set w/Remv MicroClave Clear, Sterile R, UDI: (01)00887709088139

Reason for recall

There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences.

Distribution

US:AK,AL,AR,AZ,CA,CO,CT,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,SD,TN,TX,UT,VA,WA,WI,WY OUS: Canada *** "REF" numbers AH7117 and B30211 were not marketed or distributed within the US***

Key facts

Status
Terminated
Initiation date
2019-10-04
Report date
2020-05-13
Termination date
2021-12-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Clemente, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1915-2020