FDA recall Z-1916-2024

Synthes (USA) Products LLC · Class II · device

Product

STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during orthopaedic surgery. Part Number: 03.130.010

Reason for recall

Stardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Key facts

Status
Ongoing
Initiation date
2024-04-25
Report date
2024-06-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
West Chester, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1916-2024