FDA recall Z-1921-2018

Smith & Nephew, Inc. · Class II · device

Product

smith&nephew ANTHOLOGY POROUS HIGH OFFSET STEM, 12/14 TAPER, 7H SIZE, 32 MM x 112 MM, REF 71356107 Hip prosthesis.

Reason for recall

Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm.

Distribution

US nationwide distribution.

Key facts

Status
Terminated
Initiation date
2018-05-02
Report date
2018-05-30
Termination date
2020-09-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
Memphis, TN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1921-2018