# FDA recall Z-1921-2024

> **Thoratec Corp.** · Class I · device recall initiated 2024-05-08.

## Product

Thoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386, 1286A-CAN, 1286INT and L1286INT. Left ventricular assist device

## Reason for recall

System monitor screen may display atypical behavior: Overlapping screens/buttons, Frozen screen, Distorted text, blanks, or zeroes in place of values, Unresponsive buttons where the user is unable to initiate a command, which may lead to anxiety /inconvenience, hemodynamic compromise, extended surgical time, and Pump Stop command may be selected

## Distribution

Worldwide distribution.  US nationwide including Puerto Rico,  UNITED ARAB EMIRATES, AUSTRIA, BELGIUM, BULGARIA, SWITZERLAND, CYPRUS, CZECH REPUBLIC, GERMANY, DENMARK, ESTONIA, EGYPT, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, GUADELOUPE, GREECE, IRELAND, ISRAEL, IRAN, ISLAMIC REPUBLIC OF, ITALY, KUWAIT, LEBANON, LITHUANIA, LUXEMBOURG, MACEDONIA, MARTINIQUE, NETHERLANDS, NORWAY, OMAN, POLAND, PALESTINE, PORTUGAL, QATAR, ROMANIA, SERBIA, RUSSIAN FEDERATION, SWEDEN, SLOVENIA, TUNISIA, TURKEY, SOUTH AFRICA, CANADA, AUSTRALIA, HONG KONG, MALAYSIA, SINGAPORE, SOUTH KOREA, TAIWAN, THAILAND, JAPAN, BANGLADESH, INDIA, PAKISTAN, ARGENTINA, CAYMAN ISLANDS, CHILE, COLOMBIA, COSTA RICA, MEXICO, and BRAZIL

## Key facts

- **Recall number:** Z-1921-2024
- **Recalling firm:** Thoratec Corp.
- **Classification:** Class I
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-05-08
- **Report date:** 2024-06-19

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Pleasanton, CA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1921-2024