FDA recall Z-1927-2018

GETINGE US SALES LLC · Class II · device

Product

ALM XTEN Surgical Light intended to provide visible illumination for the surgical field or for the examination of the patient. Part Numbers: ARD567801093, ARD567801094

Reason for recall

Getinge has received complaints concerning the X TEN Surgical Lights whose spring arms have broken, causing light heads to separate from the assembly. The issue relates to cracking of the spring arms at the weld seams that has the potential to develop into breaks with heavy use and aging of the equipment.

Distribution

U.S. and Gov. accts.

Key facts

Status
Terminated
Initiation date
2017-11-16
Report date
2018-05-30
Termination date
2021-07-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
Wayne, NJ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1927-2018