FDA recall Z-1933-2021

Cardinal Health Inc. · Class II · device

Product

BD UF PEN 32GX4MM 100 NANO Item Number: 4292272

Reason for recall

Due to shipping delays the devices were exposed to temperature excursions of 31.9¿F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results.

Distribution

US Nationwide distribution in the states of FL, GA, SC.

Key facts

Status
Ongoing
Initiation date
2021-03-16
Report date
2021-06-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
Dublin, OH, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1933-2021