# FDA recall Z-1938-2018

> **Ad-Tech Medical Instrument Corporation** · Class II · device recall initiated 2017-12-20.

## Product

Disposable Drill Kit which includes two drill bits, two drill stops, and two adjustment wrenches, sterile. The drill bits are 30cm long and made of stainless steel. The drill kits are supplied in a sterile state and are single-use only.     Used to drill cranial holes using a stereotactic frame.

## Reason for recall

There is a possibility that DDK2-2.4-30X Disposable Drill Kits, Lot Number 111664 208140649 contained 2.8mm drill bits from DDK2-2.8-30X Disposable Drill Kits, Lot Number 111745 208140649 and vice versa.

## Distribution

Worldwide Distribution: US (nationwide) to states of:  CA, DC, FL, MI, MA, NC, NY, WA, WI; and country of: Canada.      Distributed internationally to the United Kingdom.

## Key facts

- **Recall number:** Z-1938-2018
- **Recalling firm:** Ad-Tech Medical Instrument Corporation
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2017-12-20
- **Report date:** 2018-06-06
- **Termination date:** 2018-10-12

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Oak Creek, WI, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1938-2018