FDA recall Z-1940-2020

Cook Inc. · Class II · device

Product

Formula 418 Renal Balloon-Expandable Stent, RPN FOR418-18-80-6-12, Order Number (GPN) G56611 Balloon-expandable stent made of 316L stainless steel with a slotted tube configuration. It is pre-mounted on an over-the-wire balloon catheter, which serves as the delivery system. The stent is positioned between two radiopaque marker bands, which are located inside the balloon at the proximal and distal tapers of the balloon.

Reason for recall

The firm received complaints on five devices from one specific lot where the stent dislodged from the balloon delivery system prior to deployment. Use of affected product may lead to increased procedural time and additional intervention to remove a stent that dislodged within an unintended location.

Distribution

Domestic distribution to OH, IA, VA, and NC. Foreign distribution to Canada.

Key facts

Status
Terminated
Initiation date
2020-04-15
Report date
2020-05-20
Termination date
2024-04-15
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bloomington, IN, United States