# FDA recall Z-1944-2025

> **Dexcom, Inc.** · Class I · device recall initiated 2025-05-12.

## Product

Dexcom G7 Glucose Receiver,  SKUs (part number):  STK-AT-011(MT26403-0), STK-AT-012 (MT26403-0), STK-AT-013 (MT26403-0), STK-GT-001 (MT26403-1), STK-GT-008 (MT26403-5), STK-GT-013 (MT26403-1) , STK-GT-019 (MT26403-4) , STK-GT-100 (MT26403-2), STK-GT-109 (MT26403-2) , STK-GT-113 (MT26403-3)

## Reason for recall

Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.

## Distribution

worldwide distribution - US Nationwide and the countries of  Andorra, Argentina Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Latvia, Liechtenstein, Lithuania, Luxembourg, Malta, Netherlands, New Zealand, Norway, Oman, Poland, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom.

## Key facts

- **Recall number:** Z-1944-2025
- **Recalling firm:** Dexcom, Inc.
- **Classification:** Class I
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2025-05-12
- **Report date:** 2025-06-25

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** San Diego, CA, United States