FDA recall Z-1948-2025

Ortho-Clinical Diagnostics, Inc. · Class II · device

Product

Brand Name: Solana¿ Bordetella Complete Assay Product Name: Solana Bordetella kit Model/Catalog Number: M308 Software Version: not applicable Product Description: One sales unit (kit) contains materials for 48 tests. One kit contains 48 tubes of 1.45 mL Process Buffer and 48 Reaction Tubes. Component: No, not a component

Reason for recall

It was determined that there could be a false positive rate of up to 16.7% for the affected lots associated with the Bordetella parapertussis (BPP) results (only) of the assay.

Distribution

Please see the attached list "Consignee List_CL2025-120." Note: This file contains a specific tab, titled Shipping Summary, which contains the US and ex-US shipping summary. One (1) Distributor was affected. The other four(4) Distributors were sent the notification for information only. QuidelOrtho sent the notification to their consignees because QuidelOrtho drop-shipped this product to their 31 consignees (Included in the Medical Facility total below).

Key facts

Status
Ongoing
Initiation date
2025-03-24
Report date
2025-06-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
Rochester, NY, United States