FDA recall Z-1949-2026

Hologic, Inc · Class II · device

Product

Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035.

Reason for recall

It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.

Distribution

Worldwide distribution - US Nationwide and the countries of Asia Pacific, Europe, and the Middle East.

Key facts

Status
Ongoing
Initiation date
2026-03-23
Report date
2026-05-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Marlborough, MA, United States

Primary source