FDA recall Z-1970-2023

LumiraDx · Class II · device

Product

SteriPack Sterile Polyester Spun Swabs (25 units/Pack)- Intended for sample collection in the intended nasal area. Catalog Number: 60566RevB

Reason for recall

Fails Post-Sterilization Sterile Swabs Cantilever (Bend) Testing and may be more susceptible to breakage. If the swab breaks in the nasal cavity of a patient may cause injury or medical intervention to remove part of the swab

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2023-04-25
Report date
2023-06-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waltham, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1970-2023