# FDA recall Z-1972-2023

> **Boston Scientific Neuromodulation Corporation** · Class II · device recall initiated 2023-05-12.

## Product

Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System IFU, Superion IDS Kit IFU, Surgical Technique Manual

## Reason for recall

Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury.

## Distribution

US Nationwide Distribution: IN, NY, IL, PA, FL, SC, TX, MA, VA, AL, WI, OK, MO, MI, OH, WV, NC, UT, NH, CA, NJ, TN, NV, AZ, NE, IA, CT, ME, LA, KY, MS, MD, DC, GA, CO, KS, ID, MN, AR, HI, NM, WA, SD, OR, DE, WY, AK, and ND

## Key facts

- **Recall number:** Z-1972-2023
- **Recalling firm:** Boston Scientific Neuromodulation Corporation
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2023-05-12
- **Report date:** 2023-06-21

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Valencia, CA, United States