# FDA recall Z-1974-2023

> **QUIDEL CARDIOVASCULAR INC** · Class I · device recall initiated 2023-05-25.

## Product

Quidel Triage Cardiac Panel: REF: 97000HS, 97021HS (Troponin I), 97022HS (CK-MB and Troponin I), 97000HSEU, 97000HSEUJP, 97000HZ01, 97000QIL. Lots T13664 - T14045.

## Reason for recall

Raw material issue may cause cardiac panel products to have an average of 30% troponin signal reduction across measurement range 0.05 ng/mL to 30 ng/mL, QC testing won't detect anomaly, so an alternate method is recommended, if unavailable, use recommendations to mitigate potential patient impact, which could be missed/delayed myocardial infarction diagnosis.

## Distribution

US: TX, NC, CA, LA, SC, OR, HI, OK, AZ, AL, KS, WA, FL, VA, IA, WV, MT, UT, CO MS, NM, MA, NJ, MO AK, IL, MN, WI, IN, GA, NE, MD.  OUS: GBR, ITA, FRA, DEU, MCO, EGY, LCA, CHE, ASM, VIR, IRL

## Key facts

- **Recall number:** Z-1974-2023
- **Recalling firm:** QUIDEL CARDIOVASCULAR INC
- **Classification:** Class I
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2023-05-25
- **Report date:** 2023-07-12

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** San Diego, CA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1974-2023

## Citation

> AI Analytics. FDA recall Z-1974-2023. Retrieved 2026-08-18 from https://api.ai-analytics.org/recall/Z-1974-2023. Source: US FDA. Licensed CC0.

---