CooperSurgical, Inc. · Class II · device

Product

Advincula Delineator with Ultem Plastic Soft Cup size 3.0 cm; Part No. AD750SC-KE30 The Advincula Delineator Uterine Manipulator is a single-use disposable device designed specifically for Total Laparoscopic Hysterectomy (TLH), Laparoscopic Assisted Vaginal Hysterectomy (LAVH), and/or Laparoscopic Supra-Cervical Hysterectomy (LSCH).

Reason for recall

The soft, flexible cup of the Advincula Delineator may be unvalidated and could potentiall crack when subjected to excessive compressions during use, leading to serious health consequences, such as lacerations.

Distribution

US distribution to state of: AZ, CA, GA, IN, LA, MA, NC, NH, NJ, NY, OH, PA, VA, and WA..

Key facts

Status
Terminated
Initiation date
2019-06-14
Report date
2019-07-24
Termination date
2020-05-19
Voluntary/Mandated
Voluntary: Firm initiated
Location
Trumbull, CT, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1978-2019