FDA recall Z-1980-2019

Elekta, Inc. · Class II · device

Product

Elekta Unity, Image-Guided Radiation Therapy System Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan.

Reason for recall

The QA software solution to perform the MR to MV alignment check, does not display the stored MR to MV offset values. The user is unable to independently inspect the values during their QA.

Distribution

US Distribution to states of: TX & WI, and Internationally to: Canada, Denmark, Germany, Italy, Netherlands, Sweden, and the UK.

Key facts

Status
Terminated
Initiation date
2019-06-24
Report date
2019-07-24
Termination date
2021-04-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
Atlanta, GA, United States