FDA recall Z-1980-2023

SenTec AG · Class II · device

Product

V-Sign Sensor 2, REF VS-A/P/N, a component of Sentec Digital Monitoring System (model SDMS).

Reason for recall

The sensors may experience an out-of-box failure because after recalibration, the sensors stayed in the software mode "production" and were not reset to "released" mode.

Distribution

US Nationwide distribution in the states of CA, FL, IL, MA, MS, MT, NV, and TX.

Key facts

Status
Ongoing
Initiation date
2023-05-02
Report date
2023-06-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
Therwil, Switzerland

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1980-2023