FDA recall Z-1980-2025

Ohio Medical Corporation · Class II · device

Product

GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S

Reason for recall

Erroneous calibration values, potentially resulting in: as the device ages, oxygen purity will gradually decrease to the point where the device alarms to indicate low oxygen purity will be incorrect.

Distribution

US Nationwide distribution in the states of AR, FL, VA, ND, MS, MN, CA, NV, KY.

Key facts

Status
Ongoing
Initiation date
2025-05-08
Report date
2025-06-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Gurnee, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1980-2025