FDA recall Z-1993-2025

Diagnostica Stago, Inc. · Class II · device

Product

STA - Owren Koller. Model/Catalog Number: 00360 STA - Owren-Koller is a buffer solution intended for use as a diluent for reagents and human plasmas in coagulation tests performed with analyzers of the STA-R, STA Compact, STA Satellite and ST art families.

Reason for recall

Potential for shorter clotting times in clotting tests which have a dilution, leading to possible overestimate of around 1%

Distribution

The affected lot was distributed to customers in various US states in October and November 2024. The lot was also distributed in other countries including Canada. The lot was provided to two Distributors in US Territories: MedPharm Corp (Guam) and Isla Lab Products (Puerto Rico); MedPharm will be notified by Stago's Asia-Pacific business unit in the coming days and is not part of this report.

Key facts

Status
Ongoing
Initiation date
2025-05-21
Report date
2025-06-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Parsippany, NJ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1993-2025