FDA recall Z-1996-2021

Flowonix Medical Inc · Class II · device

Product

Physician Order Form (PL-15400-00, Nov. 2018) distributed to end user used with the following: Prometra Programmable Infusion Pump System: MODEL/CATALOG #(s): Catalog No. US Model GTIN-14 11827 Prometra Programmable 20mL Pump 00810335020082 13827 Prometra II Programmable 20mLPump 00810335020228 16827 Prometra II Programmable 40mLPump 00810335020242

Reason for recall

Physician Order Form (PL-15400-00 Nov 2018) distributed is discontinued due to the text 'Refill the pump and program refill with the same medication removed prior to MRI' may be misinterpreted as 'reuse the same drug solution extracted from the pump prior to the MRI procedure', creating a possible risk of infection

Distribution

US Nationwide distribution.

Key facts

Status
Terminated
Initiation date
2021-05-26
Report date
2021-07-07
Termination date
2024-02-20
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mansfield, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1996-2021