FDA recall Z-1997-2026

Heraeus Medical GmbH (Dental Division) · Class III · device

Product

Heraeus, PALAMIX uno. Material Number: 66057893.

Reason for recall

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

Distribution

Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;

Key facts

Status
Ongoing
Initiation date
2026-03-25
Report date
2026-05-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Wehrheim, Germany

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1997-2026