FDA recall Z-2004-2020

Inpeco S.A. · Class II · device

Product

Abbott Accelerator a3600, ARCHITECT cSystem c8000/c16000 Interface Module, Part Numbers: ACP-208-00 - Product Usage: The FlexLab (aka Accelerator a3600) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

Reason for recall

When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.

Distribution

Worldwide distribution - US Nationwide including in the states of Texas, West Virginia, South Carolina, Delaware, Louisiana, Illinois, Georgia, Florida, New York, Virginia, Arkansas, Wisconsin, California, Minnesota, Missouri, North Carolina, Utah, Michigan, Ohio, Alabama, Oklahoma, Tennessee, Kansas, Hawaii, Arizona, Washington

Key facts

Status
Terminated
Initiation date
2020-04-07
Report date
2020-05-20
Termination date
2022-07-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
Lugano, N/A, Switzerland

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2004-2020