FDA recall Z-2005-2023

Permobil · Class II · device

Product

PushTracker E3 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application

Reason for recall

When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.

Distribution

Worldwide distribution.

Key facts

Status
Ongoing
Initiation date
2023-05-10
Report date
2023-07-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Lebanon, TN, United States