FDA recall Z-2005-2026

Siemens Healthcare Diagnostics Inc · Class II · device

Product

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;

Reason for recall

Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.

Distribution

Worldwide distribution - US Nationwide and the countries of Italy, Spain, Austria, Portugal, Ireland, France, Romania, Canada, India, Philippines, Mexico.

Key facts

Status
Ongoing
Initiation date
2026-04-02
Report date
2026-05-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Norwood, MA, United States