FDA recall Z-2006-2020

Ra Medical Systems, Inc. · Class II · device

Product

Ra Medical Systems DABRA Catheter 5F (1.5mm), REF: 1222-5000-01, LOT: 18-090400-094, USE BY: 2019-09-24, MFR DATE: 2018-09-24, CE, Laser Radiation, STERILE (Gamma Sterilized), Single Use Only Revised Label: Product Label - Ra Medical Systems DABRA Catheter 5F (1.5mm), REF: 1222-5000-01, LOT: 19-092400-041, USE BY: 2019-12-13, MFR DATE: 2019-10-14, CE 2797, Laser Radiation, STERILE (Gamma Sterilized), Single Use Only Package Label - DABRA 101 5 French, Ref: 1222-5000-01, Lot: LOT: 19-092400-041, USE BY: 2019-12-13, UDI's (01)00842362100031, (17)20191213(21)2019092400041

Reason for recall

The firm has become aware there is a potential problem with its DABRA catheters resulting in higher catheter calibration failure rate due to the product shelf life is currently less than the labeled shelf life.

Distribution

Worldwide Distribution - US Nationwide - MI, TN, CA, FL, OR, PA, AZ, AL, VA, GA, TX, ID, AR, NC, IN, SC, CT, NY, KS, LA, NV, MS, NE, MD, IL, OK, MA, WA, MO, WI, KY. OUS - Netherlands, Pakistan, United Kingdom, Italy, France

Key facts

Status
Terminated
Initiation date
2019-09-19
Report date
2020-05-20
Termination date
2023-07-31
Voluntary/Mandated
Voluntary: Firm initiated
Location
Carlsbad, CA, United States

Primary source