FDA recall Z-2006-2026

Elekta, Inc. · Class II · device

Product

Brand Name: Leksell Vantage Arc System REF: 1053958

Reason for recall

Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.

Distribution

US: AR AZ CA CO DC FL GA ID IL KS MA MD ME MO MT NC NE NH NJ NM NV NY OH OK PA SC TX UT VA WA WI WV OUS: Algeria Australia Austria Azerbaijan Belgium China Colombia Costa Rica Croatia Czech Republic Egypt Estonia Finland France Germany Greece Hong Kong Hungary India Israel Italy Japan Korea, Republic of Kuwait Lithuania Mexico Morocco Netherlands New Zealand Norway Pakistan Panama Poland Portugal Saudi Arabia Singapore Slovenia South Africa Spain Sweden Switzerland Taiwan Turkey United Kingdom

Key facts

Status
Ongoing
Initiation date
2026-03-19
Report date
2026-05-06
Voluntary/Mandated
Voluntary: Firm initiated