FDA recall Z-2024-2020

K2M, Inc · Class II · device

Product

K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90157, UDI # 10888857108707, Size 28x12mm, 6 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.

Reason for recall

The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.

Distribution

Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.

Key facts

Status
Terminated
Initiation date
2020-04-13
Report date
2020-05-27
Termination date
2021-11-15
Voluntary/Mandated
Voluntary: Firm initiated
Location
Leesburg, VA, United States