FDA recall Z-2027-2026

Medline Industries, LP · Class II · device

Product

Medline medical procedure kits, containing Medline Neuro Sponges (1050 in total), BIOBURDEN TEST KIT, NEURO, ROBOTIC, ACDF, ENT, etc.

Reason for recall

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

Distribution

Worldwide distribution.

Key facts

Status
Ongoing
Initiation date
2026-03-19
Report date
2026-05-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2027-2026